Customization: | Available |
---|---|
Customized: | Customized |
Product Name: | Auto Cbc Machine Blood Hematology Analyzer |
Still deciding? Get samples of US$ 2740/Piece
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Auto CBC machine blood hematology analyzer
Technical Parameters of Auto CBC machine blood hematology analyzer
Working Mode: | Automatic sampling,detecting automatic testing, furnished with dilutor,no need diluting outside. Automatic dispensing lyse and making second diluting automaticaly |
Block-clearing: | Soak and washing back and forth,high pressure burning. Boast the function of monitoring incomplete blockage |
Alarm: | Automated system diagnosis,self testing, fault alarm and reagent shortage alarm |
Automated Mainten | Automated washing during power on and power off, reducing the blockage rate |
Standby Protectior | 24 hours switch on,automatic sleep and wake up function |
Storage: | Large quantity test results storage,at least 400 thousand test results with comprehensive information. Extended sD memory supported |
Control Mode: | Quality control mode of L-J,x-B and X-r.Westguard quality control regulation |
Calibration Mode: | Manual and automatic calibration available.Fresh blood calibration available also |
Reference Value Sett | Many reference value ranges (children, adult and the aged).The appropriate reference value range could be automatically selected according to the inputted age and sex. |
Discriminator Func | Float and fix discriminator.Manual adjust available |
Display: | TFTLCD(Optional Touch Screen) |
Input Mode: | Standard mouse and keyboard.Optional bar code reader |
Print Mode: | Internal high-speed thermal printer.Optional connection with specified external printer.English printouts |
Report Format: | Many kinds of report format (not less than 8 kinds) |
Reagents System: | Open reagent system.No cyanogen reagents.Provide original reagents. |
Connection Port: | USB,RS-232,network interface,LPT |
Dimension: | 345mm×245mmx330mm |
Weight: | 12Kg |
Economical
|
|
Fast & Convenient
|
Automatic sleep and wake up,saving energy.The sampling needle be automatically back and out while stand by and wake up.It protects safe sampling and avoids contamination caused by long time exposure. Large quantity test results storage,at least 400 thousand test results with comprehensive information.Extended SD memory supported. Many reference value ranges (children,adult and the aged). The appropriate reference value range could be automatically selected according to the inputted age and sex. | |
Accurate AnalysisPerfect Quality Analysis mode includes venous blood, capillary blood and prediluted. Quality control mode of L-J,x-B and X-r. Westguard quality control regulation. Manual and automatic calibration available.Fresh blood calibration available also. Furnished with original reagents.Conform to the demand of International Standard Traceable System. |
|
Convenient Maintenance Safety &Eco-protection Automatic & Intelligent function of diagnosis and maintenance;the log sheet for maintenance and malfunction could be established automatically. Soak, back & forth flush, high pressure burning block-clearing. Leaf door construct for convenient maintenance. |
|
|
Unit:mL |
|||
Item |
frequence |
time |
diluent |
detergent |
lyse |
Start UP |
day |
1 |
106.1 |
16 |
6 |
Shut down |
day |
1 |
40 |
22 |
|
Analysis Action |
day |
150 |
16 |
1 |
1 |
Flush |
month |
0 |
1 |
|
|
En-Flush |
month |
0 |
37.8 |
|
|
Clean |
day |
0 |
8 |
4 |
|
En-clean |
month |
0 |
32.2 |
9 |
|
Maintain |
month |
0 |
70 |
9 |
|
calibration with blood control |
month |
0 |
16.7 |
1 |
0.84 |
Blank test |
day |
0 |
16.7 |
1 |
0.84 |